What is it about?
Results of platelet counting in blood may be different, due to analyser technology. In patients with very low platelet counts these differences can lead to inappropriate decisions regarding platelet transfusion. We compared two different analyzers and a reference platelet counting method and showed that platalet transfusions may be withheld from patients due to incorrect counting results.
Featured Image
Why is it important?
Our findings give insight into differences due to analyser technology in two frequently used hematology instruments.
Read the Original
This page is a summary of: Comparative evaluation of platelet counts in two hematology analyzers and potential effects on prophylactic platelet transfusion decisions, Transfusion, September 2018, Wiley,
DOI: 10.1111/trf.14886.
You can read the full text:
Contributors
The following have contributed to this page