What is it about?
Background: The intraosseous temperature during implant installation has never been evaluated in an in vivo controlled setup. The aims were to investigate the influence of a drilling protocol and implant surface on the intraosseous temperature during implant installation, to evaluate the influence of temperature increase on osseointegration and to calculate the heat distribution in cortical bone. Methods: Forty Brånemark implants were installed into the metatarsal bone of Finnish Dorset crossbred sheep according to two different drilling protocols (undersized/non-undersized) and two surfaces (moderately rough/turned). The intraosseous temperature was recorded, and Finite Element Model (FEM) was generated to understand the thermal behavior. Non-decalcified histology was carried out after five weeks of healing. The following osseointegration parameters were calculated: Bone-to-implant contact (BIC), Bone Area Fraction Occupancy (BAFO), and Bone Area Fraction Occupancy up to 1.5 mm (BA1.5). A multiple regression model was used to identify the influencing variables on the histomorphometric parameters. Results: The temperature was affected by the drilling protocol, while no influence was demonstrated by the implant surface. BIC was positively influenced by the undersized drilling protocol and rough surface, BAFO was negatively influenced by the temperature rise, and BA1.5 was negatively influenced by the undersized drilling protocol. FEM showed that the temperature at the implant interface might exceed the limit for bone necrosis. Conclusion: The intraosseous temperature is greatly increased by an undersized drilling protocol but not from the implant surface. The temperature increase negatively affects the bone healing in the proximity of the implant. The undersized drilling protocol for Brånemark implant systems increases the amount of bone at the interface, but it negatively impacts the bone far from the implant.
Featured Image
Why is it important?
Within the limitations of the present study, it was shown that different drilling protocol for the Brånemark implant system affects both the intraosseous temperature during implant installation and the peri-implant bone healing. The undersized drilling protocol provokes a greater increase of bone temperature in the proximity of the implant compared with non-undersized drilling. The temperature at the bone-implant interface may exceed the critical value for thermal necrosis, and it may have negative effects on peri-implant bone healing. The present results indicate that undersized drilling increases the amount of bone in the proximity of the implant, but it has a negative impact on the bone area far from the implant surface. The moderately rough surface does not influence the bone temperature, while it increased the bone attached to the implant. Further studies are needed to confirm the present results and to deeply investigate the thermal behavior and the biologic effect of peri-implant bone overheating during implant installation and to provide guidelines on the clinical decisions for the proper drilling protocol, based on the bone quality and implant design.
Perspectives
Read the Original
This page is a summary of: Intraosseous Temperature Change during Installation of Dental Implants with Two Different Surfaces and Different Drilling Protocols: An In Vivo Study in Sheep, Journal of Clinical Medicine, August 2019, MDPI AG,
DOI: 10.3390/jcm8081198.
You can read the full text:
Resources
Contributors
The following have contributed to this page