What is it about?
This is one of the four studies on which ExplantLab collaborated with FDA listed below: • The clinical implications of metal debris release from the taper junctions and bearing surfaces of metal-on-metal hip arthroplasty: joint fluid and blood metal ion concentrations • Shorter, rough trunnion surfaces are associated with higher taper wear rates than longer, smooth trunnion surfaces in a contemporary large head metal-on-metal total hip arthroplasty system • A comparison study of stem taper material loss at similar and mixed metal head-neck taper junctions • Material loss at the femoral head taper This group of papers showed: • CoCr head tapers were associated with greater volumetric wear if mated with a rough (microgrooved) stem taper when compared to those mated with smooth male tapers. • When mated with CoCr femoral heads, CoCr femoral stems lose proportionately more material through tribocorrosion than Ti femoral stems. • Well functioning tapers release almost negligible amounts of metal in volumetric terms. • Head diameter is a crucial factor in determining taper dysfunction. • Failing taper junctions release chromium debris which is preferentially taken up into red blood cells, implying that hexavalent chromium species are released.
Featured Image
Photo by Headway on Unsplash
Read the Original
This page is a summary of: The clinical implications of metal debris release from the taper junctions and bearing surfaces of metal-on-metal hip arthroplasty: joint fluid and blood metal ion concentrations, The Bone & Joint Journal, June 2016, Bone & Joint,