What is it about?
An in-house trajectory log analysis program (LOGQA) was developed to evaluate the delivery accuracy of volumetric-modulated arc therapy (VMAT) for stereotactic body radiation therapy (SBRT). Methods have been established in LOGQA to provide analysis on dose indices, gantry angles, and multi-leaf collimator (MLC) positions. Three types of intentional errors were introduced to patient plan for the differentiation of log file-based quality assurance (QA) from traditional phantom measurement.
Featured Image
Photo by National Cancer Institute on Unsplash
Why is it important?
Machine log analysis provides crucial information on VMAT SBRT plan delivery. End-to-end plan transfer and beam parameters accuracy check using log files provides a robust and reliable QA method to reduce patient-specific QA measurement, provided that a comprehensive measurement-based machine-specific QA program is in place.
Perspectives
Read the Original
This page is a summary of: Patient‐specific quality assurance using machine log files analysis for stereotactic body radiation therapy (SBRT), Journal of Applied Clinical Medical Physics, October 2020, Wiley,
DOI: 10.1002/acm2.13053.